Replacement Parts Industries, Inc.

OEMs & Models > AUTOCLAVES AND STERILIZERS > SCICAN > Statim 2000 > SCK020

AUTOCLAVES AND STERILIZERS
SCICAN
Statim 2000

Tech Support Library

RPI Part #SCK020
OEM Part #01-100560S

AIR COMPRESSOR KIT

RPI Part #SCK020 - AIR COMPRESSOR KIT
  • 28mA @ 115VAC; 50/60 HZ.
  • Rated air flow: 0.177 cfm
  • Rated pressure: 2.84 PSI
  • Wire leads: #22 AWG x 22” long
  • See Notes: below

Fits: Chassis
Model(s) This Part Fits: Click here for listing.

Catalog Page(s): 10-1504

Part Description Updated: 2/3/2024



Notes:

Service Tip #1: A leaky check valve can cause moisture or steam to seep inside a compressor thus causing the compressor to fail. To avoid this problem, replace the check valve when replacing the compressor and replace the check valve every 2 years.

Service Tip #2: - The OEM recommends replacing the Compressor Filters every 6 months or every 500 cycles (whichever comes first).

The RPI Advantage - The RPI tubing in the Check Valve Kit (RPI Part #SCK011) is translucent - making it easy to see a leaky check valve. For your convenience, RPI includes a check valve in the Air Compressor Kit (RPI Part #SCK020)!

Includes:

TO DISPLAY PRICING, RPI CUSTOMERS CLICK HERE TO LOG IN.
(All others please call or fax for pricing.)

Click here to list all parts for SCICAN Statim 2000

Format this page for printing

Home Back