OEMs & Models > AUTOCLAVES AND STERILIZERS > SCICAN > Statim G4 2000 > SCK020 AUTOCLAVES AND STERILIZERS
| |
RPI Part #SCK020
Fits: Chassis
Part Description Updated: 2/3/2024 |
Service Tip #1: A leaky check valve can cause moisture or steam to seep inside a compressor thus causing the compressor to fail. To avoid this problem, replace the check valve when replacing the compressor and replace the check valve every 2 years. Service Tip #2: - The OEM recommends replacing the Compressor Filters every 6 months or every 500 cycles (whichever comes first). The RPI Advantage - The RPI tubing in the Check Valve Kit (RPI Part #SCK011) is translucent - making it easy to see a leaky check valve. For your convenience, RPI includes a check valve in the Air Compressor Kit (RPI Part #SCK020)! Includes:
|